Q   How many data points can I log with the DaqPRO?

A   New Page 2

Sampling points:

Increasing the number of active inputs limits the number of sampling points one can choose. The following condition must be always satisfied:

Number of samples ´ Active Inputs £ Memory

DaqPROs memory is sufficient for 512,000 samples.

The table below lists the maximum number of sampling point according to the number of sensors:

Number of sensors

Maximum number of sampling points

1 512,000 (exact number: 522,752)
2 256,000 (exact number: 260,608)
3 128,000 (exact number: 129,536)
4 128,000 (exact number: 129,536)
5 64,000
6 64,000
7 64,000
8 64,000

 

Continuous sampling

·          Continuous sampling is possible up to a maximum sampling rate of 100/s.

·          The data must be presented in a display mode other than graphic.